| User's Manual for DRAGON Separator Hardware 
Revision history
 
 
| Version | Author | Description | Date |  
| 1.0 | D.A. Hutcheon | Initial document | 3 January 2003 |  
| 1.1 | D.A. Hutcheon | Details about FCM2 | 2 April 2003 |  
| 1.2 | D.A. Hutcheon | ED DAC fix | 9 June 2003 |  
| 1.3 | D.A. Hutcheon | FCM2/Beta-1 switch | 20 Nov 2003 |  
| 1.4 | D.A. Hutcheon | FCM2/AUTO roughing | 20 Mar 2007 |  
| 1.5 | D.A. Hutcheon | Beta Mon effcy note | 23 June 2008 |  | 2.0 | D.A.Hutcheon | Decadal update | 20 March 2019 |  Contents
Magnets
    
    Dipoles
    Quadrupoles
    Sextupoles
    Steerers
    Electrostatic dipoles
Diagnostics
    
    Faraday cups
        
	FCF and the MCP
        FCM2
	Slits
    Beam centering monitors
    Beta monitor
    CCD camera
    Vacuum
    
    Pumping down a slit box volume
    Venting a slit box section to 1 atm
    The ED tanks
        
	Recovery after a power outage
	Safety fence
Where is that controller?
 The DRAGON magnets include 2 dipoles (MD1, MD2), 10 quadrupoles (Q1-Q10), 
 4 sextupoles (SX1-SX4) and 5 double steerers (SM0-SM4).  Regulated DC 
 power supplies for the magnets are located on or under the (purple) DRAGON 
 platform.  The power supplies are interlocked to thermal switches (all 
 magnets) and to flowmeters for cooling water (all except SM1 and SM2).
 
 The flow of cooling water to each magnet is monitored by a meter connected 
 to 
 an "intelligent flow controller" box mounted at the east railing on the 
 upper platform.  Table 1 gives the channel 
 assignments in the flowmeter controllers.  Water flow and 
 thermal switches attached to the magnet coils are interlock inputs to the 
 magnet power supplies.  If EPICS indicates no water flow, check first to 
 see if the rotor of the flowmeter (located at the magnet) is turning. 
 If the rotor is turning rapidly, go up to the flow controller, open the 
 front cover and look for a flashing light, indicating good flow, on the 
 channel in question.  If the rotor is turning but the interlock condition 
 cannot be reset (yellow text changing to black) by the RST button , 
 consult 
 Controls group.  If the rotor is not turning, either the water has been 
 turned off or the flowmeter has stopped working: call Beamlines group or 
 consult with ISAC Operations. 
 
Table 1:  Channel assignments in DRAGON flow controllers. 
 
Designed to bend particles with maximum rigidity of 0.5 Tesla-m (150 
 MeV/c), the dipoles have a maximum rated current of 500 A.   
 (See G.S. Clark design notes TRI-DN-98-12 and -98-10.) 
 The magnet power supplies are designed to give a DC current stable to 
 0.01%. The MD1 power 
 supply is beneath the platform on the north side, and the MD2 supply to 
 the south.  The AC breakers for the Philtek power supplies are located in 
 a panel under the platform at the foot of the stairway.
| Unit | Channel | Magnet |  
| Upper | 1 | Q1 |  
| 2 | Q2 |  
| 3 | Q3 |  
| 4 | Q4 |  
| 5 | Q5 |  
| 6 | Q6 |  
| 7 | Q7 |  
| 8 | Q8 |  
| Middle | 1 | Q9 |  
| 2 | Q10 |  
| 3 | MD1 |  
| 4 | MD2 |  
| 5 | SM0 |  
| 6 | -- |  
| 7 | SM3 |  
| Lower | 1 | SM4 |  
| 2 | -- |  
| 3 | SX1 |  
| 4 | SX2 |  
| 5 | SX3 |  
| 6 | SX4 |  
| 7 | -- |  
| 8 | -- |  
 Control of the dipole power supplies is via interface cards on a Canbus 
 circuit.  The EPICS page for MD1 (or MD2) is called from DRAGON 
 menu OPTICS | Optics(1) ( Optics(3) for MD2) and clicking on the MD 
 icon.  The page has buttons to turn the magnet On ("1") or Off ("0") and 
 to reset latched interlocks ("RST").  If the magnet refuses to turn On 
 remotely, try going to EPICS page Diagnostics | Canbus 
 diagnostics | DRA:MD1 and hit "RST P/S".   
  A Setpoint slider and current readback are calibrated in Amperes. 
 
 The MD magnetic field is sensed by an NMR probe (#3 probe for MD1 and 
 #4 probe for MD2) inserted through a hole in the "inside" return yoke 
 at beam height.  A jig holds the probe at a well-defined position and 
 orientation, just outside the vacuum vessel, in a region where the field 
 is still flat.  Each probe is connected to a CERN-type controller located 
 in the electronics rack under the platform.  Controller output (field in 
 Gauss) is displayed on a front-panel display and is sent also to EPICS 
 where it is presented as part of the MD1 (or MD2) panel.  The controller 
 can search and lock on the resonance - when it locks, the field value 
 displayed by EPICS changes from white to blue.  The (rather narrow) search 
 range is set by a 0-6.5V level fed in the back of the controller unit. 
 The level comes from a supply which is controlled by slider/On/Off/status 
 buttons on the bottom row of the MD1 entry of Optics(1) (Optics(3) for 
 MD2).
  
 The search range is not quite a linear function of input voltage (slightly 
 S-shaped): if the NMR is not locked, observe the search range of the 
 displayed (white) field values and adjust the voltage to make the search 
 range bracket the expected field.  (The expected value for MD1 can be 
 calculated from the Setpoint current: B = 11.7*I + 100.  For MD2: 
 B = 16.5*I.) 
Quadrupoles 1-8 were built to TRIUMF design by Sunrise Engineering 
 (Delta).  Two (Q1, Q6) are `"4-inch" quads (see G.M. Stinson 
 design note TRI-DNA-98-4); five (Q3,Q4,Q5,Q7,Q8) are "6-inch" quads (see 
 G.M. Stinson, TRI-DNA-98-5); one (Q2) is "6-inch" with a 6% sextupole 
 component (G.M. Stinson, TRI-DNA-99-3).  The 4-inch quads were designed to 
 produce a field gradient of 0.5 kG/cm at maximum current 325 A, the 6-inch 
 quads 0.36 kG/cm at 325 A. 
 Quads Q9 and Q10 (CERN "Smit-Elma") have a 15-cm aperture (see 
 G.S.  Clark design note TRI-DN-98-17).  Pole tip fields were surveyed to 
 be 19 Gauss/Amp.   
 
 The quadrupole power supplies, designed for 0.1% current stability, are 
 rack-mounted units (Xantrex and Power Ten) 
 located on the power-supply platform. 
 Their AC breakers are in panel P456 at the east railing of the platform. 
 
 Quadrupole fields are measured by Hall probes, jig-mounted in the middle 
 (longitudinally) of each quad.  The probes for Q1-Q8 are located in the 
 gap between poles, at the radius where the field is a maximum.  On Q9 and 
 Q10 this location was not accessible.  The Hall probe control boxes, 
 located close to the quads, are on a Canbus daisy-chain.  When viewing 
 quadrupole field values on the EPICS Optics pages, make certain that the 
 page is wide enough so that the high-order digits are displayed. 
 Expected B/I readings for Q1 and Q6 are 11.6 Gauss/Amp; for Q3,4,5,7,8 it 
 is 12.5 Gauss/Amp; for Q2, 12.9 Gauss/Amp; for Q9 and Q10, 
 nnn Gauss/Amp. 
SX1 and SX2, 6-inch sextupoles originally installed in meson channel M13, 
 were surveyed to have pole-tip fields 18.5 Gauss/Amp.  SX3 and SX4 came 
 from channel M15; they give 18.1 Gauss/Amp. 
 The sextupole power supplies are rack-mounted units (Xantrex) 
 located on the power-supply platform. 
 Their AC breakers are in panel P456 at the east railing of the platform. 
 
 They are not equipped with Hall probes. 
The double-steerers can deflect ions of maximum rigidity (0.5 T-m) by up 
 to 25 mrad in x and/or y.  Steering magnets SM1 and SM2, obtained from 
 Chalk River, are air-cooled and have maximum rated current of 5 A.   
 (If they are to be used above 3 A, a fan or other forced-air cooling 
 should be installed.)  SM3 and SM4 are water-cooled, with maximum current 
 100 A and have 6-inch apertures (see G.M. Stinson design note 
 TRI-DNA-99-1); they have 3.88 Gauss/Amp.  SM0 (a.k.a. "the Wobbler") is 
 used for separator optics studies, for which it is mounted at the target 
 position.  It is a 4-inch design (G.M. Stinson, TRI-DNA-98-7), water 
 cooled, maximum current 100 A, having 3.24 Gauss/Amp. 
 The steerer power supplies are rack-mounted units (Xantrex and zzz) 
 located on the power-supply platform. 
 Their AC breakers are in panel P456 at the east railing of the platform. 
 When SM0 is not in use (most of the time), its power supply will be locked 
 out; thus, most of the time the EPICS icons for SM0X and SM0Y are red. 
Dipoles ED1 and ED2 consist of polished cylindrical titanium electrodes 
 having a gap 
 of 100 mm and radii of curvature 2 m for ED1, 2.5 m for ED2.  Design 
 bending power is 8 MV (energy/charge 4 MeV/q) which requires electrode 
 voltages of ±200 kV on ED1, ±160 kV on ED2.  They are housed in 
 large cylindrical tanks.  The HV units are stacks powered by Glassman 
 r.f. supplies; the stacks are immersed in SF6 gas at 2 atm, enclosed in 
 re-entrant ceramic insulating cylinders. 
 The tank is encased in 6 mm of lead, to absorb x-rays given off during 
 voltage conditioning.  X-ray production is monitored by thick plastic 
 scintillators connected to photo-multiplier tubes, mounted above a viewing 
 port on the lid of each vacuum tank.  The PMT high-voltage supply is a 
 LeCroy unit, manually controlled, in the 19'' rack below the platform. 
 Customary HV setting is -2000 V.  A discriminator in the same rack 
 provides logic signals which are made available to the EPICS control system.
 
 The Glassman supplies are close to their respective electrodes, in a 
 19'' rack for the anodes and slung from the vacuum tank for the 
 cathodes.  They are controlled via a Canbus link.  The AC to the Glassman 
 supplies is interlocked: cages around the feed-throughs to the stacks must 
 both be closed; there must be less than 0.5 atm in the tank as measured by 
 a mechanical gauge; there must be less than 10-5 Torr in the tank as 
 measured by an ion gauge; at least one of the tank pumps (turbo, cryo or 
 ion) must be on.  The interlock boxes are located in the racks that hold 
 the anode power supplies. 
 
 The EPICS display shows set voltage and stack current for anode and 
 cathode.  Normally the current reading is about 1 µA per 6 kV. 
 
 Following a power glitch or a re-boot of EPICS it is not uncommon for 
 the HV to refuse to come on.  If it is not due to a genuine interlock 
 fault, go to the EPICS ED page and bring up the Expert panel.  Change the 
 "I+" setpoint to 0 then to 40, and do the same for the "I-" setpoint. 
 ("40" is the current limit in µA.) 
 If HV still doesn't come on, call a high-voltage expert. 
 
 A spark in an ED or some other electromagnetic pulse may cause the DAC
 which sets electrode current limit to lose its calibration.  A symptom of
 this is that when the HV is turned On, the voltage and current will
 momentarily ramp up but almost immediately go back down to near zero.
 A way to check for this problem is to use the "loopback" test of the
 CANBUS system: from the DRAGON EPICS menu select Diagnostics|Canbus
 and then the ED in question.  For the problem electrode (+ or -) change
 the I button from "no LB" to put it in loopback mode, whereby the DAC
 output is fed directly back into the ADC.  Run the appropriate slider bar
 up and down and you should normally see the Readback and Setpoint values
 move up and down together.  If they do so, you have a different problem
 and should call an HV Expert.  If they do not have the same value, try to
 recalibrate the DAC: momentarily unplug the CANBUS daisy chain input to
 the controller box in question (ask an ISAC Operater for help with this).
 If this doesn't fix the problem, call an HV Expert.  
 NOTE: unplugging the CANBUS will cause all CANBUS-controlled devices
 (magnets, ED) to forget their settings and they must be turned on again
 and set to proper values. 
 
 The electrodes must be "conditioned" for stable operation at high 
 voltage.  As the voltage is increased, at some point the current will jump 
 up to the current limit of the supply, vacuum will get worse by an order 
 of magnitude, and the x-ray counts will jump from from tens or hundreds 
 per second to many tens of thousands per second.  If the voltage is set 
 just above the onset of these effects, the system should return to good 
 vacuum, low x-ray counts and 1 µA per kV within a minute or so. 
 Then the voltage can be raised slightly and the cycle repeated.  For 
 detailed instructions on HV conditioning, consult the 
 EPICS routine for conditioning ED1, ED2 at high voltage.
 
 Note: If signs of conditioning persist for more than approx. 1 minute after a 
 small increase in voltage (say 10 V), this may indicate a problem such as 
 carbon tracking along the ceramic insulator or within the HV stack;
 reduce the voltage and consult an HV expert. 
DRAGON has 4 controllable Faraday cups: FC1 is between Q2 and MD1; FCCH is 
 after the 
 slits at the Charge Selection focus following MD1; FCM is after the slits 
 at the Mass Selection focus following ED1; FCF is after the Final slits at 
 the end of the separator.  They can be inserted or retracted by a 
 pneumatic actuator, controlled via EPICS.  The EPICS control panel for 
 each cup allows selection of a full-scale current range for the current 
 integration.  The readout noise is approximately 10 pA, setting a lower 
 limit on the beam currents which can be read reliably. 
 An integration time of 1-2 seconds has been built into the ADC readout 
 routine, so several seconds are needed for the readings to reach a stable 
 value. 
 Each Faraday cup is equipped with a bias ring at the front, to suppress 
 the escape of secondary electrons.  If the "good bias" indicator is not 
 green, click on "Bias" and set the bias to greater than 100 V on Reverse 
 polarity.   
 
 The bias power supplies and the charge-integrating ADCs are located in a 
 Controls VME crate in rack 23A on the platform. 
A fixed-position Faraday cup FCM2 has been mounted in the
Mass Slit Box, downstream of the slits and 12 cm to the "low
mass" side of the standard cup FCM.  (See the location
of the cup.)  The 12-cm offset should catch
beam of the selected charge state when ED1 is tuned for recoils
having 4/3 the mass of the beam.  The exact value of the mass ratio
must be determined by finding the value of ED1 which centres the
beam in FCM2 -- if 12 cm offset is not close enough to the 4/3 mass
ratio, there is provision for adjusting the offset (manually).
 Two pairs of motor-driven slits are located at each of 3 focus 
 points: XSLITC and YSLITC are the horizontally-moving and 
 vertically-moving slit pair at the Charge selection focus; XSLITM and 
 YSLITM are at the Mass selection focus; XSLITF and YSLITF are at the Final 
 focus. 
 Each member of a pair (e.g. the Left and Right parts of an XSLIT pair) is 
 driven by a stepping motor.  However, the user specifies the Width and 
 Position of the slit opening and EPICS control takes care of computing the 
 Left and Right (or Top and Bottom) positions.  The sign convention is that 
 of GIOS or TRANSPORT ion optics codes: positive x is to the left, looking 
 downstream in the direction of particle motion; positive y is up. 
 
 The stepping motor controllers are located in rack 23A on the platform, 
 below the VME crate.  Calibration of slit positions is done by means of a 
 microswitch at the "Out" position of each drive.  Following a power 
 failure or an EPICS re-boot, it is necessary to do a 
 "Calibrate" operation for each slit drive.  This causes the slit member 
 to be driven out to the outer (microswitch) limit, then in to the position 
 required by the Position and Width settings. 
 
 A 3rd microswitch protects against an attempt to close the slit pair 
 beyond the point of contact ("negative Widths").  This means that care 
 must be taken if very small Widths are called for: a Calibrate operation 
 may never complete when Width=0 if the collision microswitch engages 
 before the motor has stepped to the computed 0 width; a change in Position 
 when Width is < 2mm may produce the "collision" condition during the 
 movement of the slit members.  In the latter case, the EPICS driver will 
 stop the "pursuing" member but allow the "leader" to continue to its 
 proper destination - the indication that this has happened is that 
 the read-back position (blue number) doesn't end up at the Setpoint 
 number for Width and Position and the "Closed limit" light is green for 
 the "leader".
  The true position of XSLITC, the slits defining horizontal position
 at the Charge Slit Box, is a key datum for calculation of beam energy.
 In addition to counting step commands given to the stepping motors,
 SXLITC has position readout by a Mitutoyo linear slide, whose values
 appear in the EPICS page DRA:XSLITC as "linear scales".  (There is a
 0.3 mm offset between this scale and the true zero as determined
 by sighting through MD1.) 
 In case there is concern that the slit drive has not stepped to the 
 desired position, the X slit pairs are equipped with pointers and 
 millimeter scales to allow independent verification of the actuator 
 position.  The scales are marked also with the readings that correspond to 
 the out-limit microswitches; these numbers should appear in the EPICS 
 table used in Calibration. 
 
 As for the Faraday cups, slit currents are sent to an ADC.  The slits have 
 no bias for secondary electron suppression. 
Six beam centring monitors (BCMs), together with the 3 sets of slits and 
 the target aperture, allow beam position to be defined at two locations in 
 each of the 5 straight-line segments of the separator.  Each BCM consists 
 of 4 plates arranged in a 2x2 array, mounted on insulators and separated 
 from each other by a small gap.  The current from each plate is read by an 
 ADC and EPICS combines the 4 readings to show total current plus asymmetry 
 of current in each of left-right and top-bottom directions. 
 Insertion and retraction is by pneumatic actuator.  If the device fails to 
 reach its In limit (defined by a microswitch) in time (e.g. low compressed 
 air pressure), a Timeout is indicated.   Timeout flag may be cleared with 
 the RST button. 
 
 The current range scale must be chosen so that none of the quadrants is 
 shown with full-scale current, or else the asymmetry calculations will not 
 be valid.  There have been occasional problems with a quadrant of a BCM 
 reading a substantial current, even with no beam on it.  This may be due 
 to a flexible lead shorting out, for example, which requires venting a 
 section of the separator in order to investigate further.  A temporary 
 work-around is to steer the beam so that none of it falls on the 
 "bad" quadrant and adjust the horizontal or vertical steering for equal 
 currents on each of the two pairs of adjacent plates, in turn. 
The Beta monitor is a pair of plastic scintillators which detect beta 
 particles emitted by radioactive beam atoms which stop in the Mass slit. 
 Located in air in a well in the Mass focus diagnostics box, the Beta 
 monitor subtends about 1/1500 of 4 pi solid angle at XSLITM.  The rate 
 for coincidence detection of a particle in the two scintillators is 
 presented to the EPICS virtual scaler and the MIDAS DAQ. 
 The coincidence response depends upon the energy distribution of the decay 
 betas: a detected beta must have eneough energy to pass through the 
 entrance window, through the first scintillator (6mm) and far enough into 
 the second scintillator to be above discriminator threshold. 
 As well as the coincidence rate, the front-counter singles rate is sent to 
 the MIDAS scaler. 
Connection is by a coax cable as indicated by the blue line:   
 The HV for the plastic scintillator photo-multiplier tubes is from 
 channels 2 and 3 of the same LeCroy HV unit which supplies the ED x-ray 
 monitors.  The discriminators, coincidence unit, and level adapter modules 
 are in the same NIM bin as the NMR controllers (rack under the platform). 
 
Drawings and notes on expected efficiency for detection of betas from decay of 
Na-21 beam from DAH logbook:
  A CCD (Starlight Xpress MX7-C) views 
 upstream through an alignment port of dipole MD1, to the target position. 
 When beam is passing through a gas target, it emits light which can be 
 imaged on the CCD.  The position of the beam can be determined relative to 
 the apertures of the gas cell with sub-millimeter accuracy.  The width of 
 the beamspot can be seen also (but it is a measure of beam width only if 
 the light comes from beam ions, not target gas which may have diffused 
 before recombination). 
 The MX7-C is controlled from a dedicated PC via the Starlight Xpress 
 driver.  2-d density plots may be made (generally need to start with a 
 Linear Stretch of the intensity scale in order to see the beamspot). 
 A Photometry option gives counts for individual pixels.  The counts from 
 the entire pixel array may be stored in a .FITS file for later analysis. 
 
 The location of the gas cell aperture may be determined by closing 
 isolation valve HEBT2:IV8 and turning on ion gauge IGU3 (and opening valve 
 IV11!).   The outline of the 6mm cell entrance aperture appears backlit 
 by light scattered off IV8. 
The separator vacuum system has 5 sections, each with independent 
 high-vacuum pumps, which are connected in a common system when ions pass 
 through.  These sections are: the ED1 tank; the ED2 tank; the Charge slit 
 box, MD1 vacuum box and pipe between the Charge box and ED1 tank; the Mass 
 slit box, MD2 box and other pipe between ED1 and ED2 tanks; the Final slit 
 box and pipe between it and ED2. 
 A single roughing pump connects to one (at a time) of these volumes.  It 
 is a Leybold Dri-pump - oil-free, pumping down to about 50 mTorr. 
 Similarly, all the turbo pump backing is done by a second Dri-pump. 
 Interlock conditions on opening roughing valves (RV) or backing valves 
 (BV) protect against gross Operator errors, such as trying to rough down a 
 volume when its vent valve is open.  (However, this is true only if the 
 interlocks have not been Bypassed - if a valve has "Interlocks 
 bypassed" and you don't know why, find out why before opening it.) 
 
Turbo pumps  have a Full speed (38k rpm) and a Low speed (25k rpm) 
mode.  If, for example, a turbo pump is being run simply to maintain good 
vacuum in a slit box for a few days, it may be desirable to run it at Low 
speed to limit wear and tear.  The EPICS control panel for each turbo pump 
has a toggle switch to change between Full and Low speed modes.  An indicator 
light ("Low Speed") on the EPICS panel is yellow if Low speed is selected, 
black if Full speed is selected.  If Full speed is selected while the 
turbo pump is at Low, EPICS may turn the pump Off.  If this happens, it 
will be necessary to have the Backing pump on and connected (see note 
below regarding AUTO mode) in order to restart the turbo pump.  Note also 
that with no gas load a turbo pump will take a long time to spin down from 
Full to Low rpm.  The "At speed" EPICS light is green only if the speed is 
above a defined threshold (usually 36k rpm), whatever the selected speed; 
to know the actual turbo speed it is necessary to go to the turbo 
controller box and examine its visual display.
   
After a finite number of hours of running, the
 Dri-pumps  will have to 
undergo an expensive maintenance.  To limit the number of Roughing pump 
hours, it  should not be left running  when not needed.  The 
running time of the Backing pump can be limited by selecting the AUTO 
feature, using the "Pumping mode" toggle switch at the right-hand side of 
the main menu for EPICS control of DRAGON.  Connected to the common 
backing line is a storage tank.  In AUTO mode the Backing pump (BP31) 
remains off and isolated from the backing line, until the pressure in the 
backing line and tank reaches about 300mT.  At that pressure BP31 turns 
on, valve PV31 opens and the pump runs for 5 minutes before PV31 closes 
and BP31 shuts off.  Normally, BP31 should come on only once or twice per 
day when in AUTO mode.
 
When more than one section of the separator is to be pumped down, it is 
advisable to switch from AUTO to MANUAL pumping mode.  The reason may be 
found in the Device Note for PV31: "When in automode this device is forced 
off for 15 minutes after CG31A reaches less than 80 mtorr".  The backing 
pump has the hidden condition "In auto mode this device is forced off as 
long as PV31 is forced off: see PV31 note."  Attempting to pump down two 
sections of the vacuum may have the unexpected effect that in AUTO mode 
BP31 can be turned on "by hand" for the first section but not for the 
second section.
The following describes the steps in pumping down the Charge slit box 
 volume, but it applies to the Mass slit box or Final slit box with 
 suitable change in label of valves, pumps or gauges. 
  
 Close vent valve VV21. 
 Turn on roughing pump RP21 and when convention gauge CG21B goes 
 below 100 mTorr, open pump valve PV21. 
 Open RV21 and observe box pressure on CG21.  Pressure should 
 decrease steadily until CG21 reaches 100 mTorr, when RV21 will close. 
 If backing pump BP31 is not already running and PV31 open, turn on 
 BP31 and open PV31. 
 Open BV21 and turn on turbo pump TP21 (it may be necessary to RST 
 latched interlock conditions).  TP21 icon should turn dark green and 
 CG21 drop to a near-zero reading.  After 10-15 minutes the turbo should 
 get  up to speed and its icon change from dark green to light green. 
 Turn on ion gauge IG21.  If the volume has been at atmosphere, 
 initial IG21 readings may be in the 10-5 Torr range and will only 
 slowly (over hours) decrease to 1-2 × 10-6 Torr as water is 
 desorbed from inside surfaces. 
  
 The Final slit box has the complication that it may contain fragile MCP 
 target foils.  Roughing down begins through RV54, which is in series with 
 a manual throttle valve that has been adjusted to limit initial pumpdown 
 to a foil-friendly rate.  If it is known for certain that no foils are 
 mounted (e.g. by looking in the window of the MCP Foil mounting plate), 
 fast pumpdown  may be done through RV52 (with Bypassed interelocks).
 
The Final slit box is followed either by a small vacuum box containing a 
Si strip detector or by a gas-filled ionization chamber.  In the first 
case the "ED2 to SSD" option should be selected under the Vacuum menu.
For the ionization chamber choose "ED2 to IC".  Operation of the 
ionization chamber is complicated and its use of the vacuum system is 
detailed in the section describing general operation of the IC.  
This procedure for the Charge slit box applies also to the Mass and Final 
 slit boxes, with suitable change of labels on gauges, pumps and valves. 
  
 Only rarely will the ED tanks be vented to atmosphere.  The general user 
 is strongly discouraged from doing this, because of the lengthy time it 
 takes to recover high vacuum (10-7 Torr level) after a venting. 
 More likely will be a requirement to resume pumping if all the pumps 
 are off following a power outage.  Each tank is equipped with:Close isolation valves IV11 and IV21. 
 Turn off ion gauge IG21 and turbo TP21, and close backing valve BV21 
 \item toggle vent valve VV21 open and closed as quickly as possible, to 
 start the turbo braking process.  CG21C may climb to approx. 10 Torr, and 
 then gradually decrease as TP21 spins down.  When CG21C reaches 1-2 Torr, 
 allow another quick gulp through VV21. 
 When CG21C and CG21 become nearly equal, start the main venting: 
 at the compressed Nitrogen cylinder beside MD2, verify that the manual 
 valve "LN2 dewar pressure" is closed and "Separator vent" is 
 open.  Set the "dead-man switch" timer to 15 minutes. (In case you get
distracted midway through the venting process, this prevents venting the 
whole cylinder of nitrogen by mistake.)
Locate the vent line pressure relief valve (downstream of MD2, knee 
 height, by valve VV21A.  Adjust the pressure regulator (clockwise 
 increases pressure!) until gas can be felt escaping the pressure relief 
 valve.   
 Open VV21 and watch pressure rise on CG21.  When it reaches 760 
 Torr, valve off the nitrogen cylinder. 
 
 A turbo pump (Varian 1000 l/s) used to pump down from 
 "intermediate" vacuum (too high for the ion gauge, too low for a 
 convection gauge) to the 10-6 Torr range.  Otherwise, usually turned 
 off and valved off to reduce wear \& tear. 
 A cryo pump used for its high pumping speed, during DRAGON operation 
 with beam.  A single compressor runs 2 cold-heads, one at each ED tank. 
 It is located near the Mass slit box, next to the Roughing and Backing 
 pumps. 
 An ion pump used to maintain vacuum when DRAGON is not in use for 
 extended periods.  It must be turned on or off manually - only the 
 pressure readout is sent to EPICS, to indicate when the ion pump is on. 
  
 At least one of the 3 pumps must be on in order to turn on the tank ion 
 gauge and (because they are interlocked to the IG) the electrode high 
 voltage supplies. 
The interlocks on 
  the roughing valves (RV33, RV53) are such that they allow 
 only the pump-out of a warm (> 250 K) head, and the cryos (CP33, 
 CP53) can be turned on only if they have been pumped down below 150 mTorr. 
 If a power outage occurs during a run, it may be necessary to pump on the 
 cryo head(s) if they have warmed up enough to release the gunk that they 
 have pumped out of the tank. 
 If the outage was brief and the pressure in 
 both CG33C and CG53C is below 100 mTorr, restart the cryos CP33 and CP53. 
 The temperatures of the CPs and the pressure at the associated CGs should 
 drop.  There should be a "rrr ... rrr ... rrr ..." sound from the 
 cold-heads at the ED tanks; if there isn't, try turning the power off and 
 on at the compressor (inside the fence, by the roughing and backing pumps) 
 and re-starting CP33, CP53.  If this doesn't work, consult an expert. 
 If the outage was long enough that CG33C or CG53C pressure rose 
 above 100 mTorr but the temperature at CP33 or CP53 stayed well below 
 250K, it will be necessary to pump away some of the gas that was released 
 by the warm-up of the cryo.
 
  
 On the DRAGON EPICS menu go to Bypass/force | Bypass vacuum and 
 click On RV33 (or RV53).  The text should change from black to yellow, 
 indicating the interlock has been bypassed. 
 Turn on RV21 and open PV21.  When CG21A < CG33C (or < CG53C), 
 open RV33 (or RV53).   
 If the pressure at CG33C (or CG53C) drops very slowly and rises 
 rapidly when the RV is closed, a lot of gunk has been released and 
 pumpdown may take a considerable time.  If the tank turbo pump TP31 (or 
 TP51) is not already running, turn it on (following the instructions as 
 given above for TP21 in the Charge slit box volume).  When TP31 (or 
 TP51) gets up to speed and provided CG31 (or CG51) is < 100 mTorr, open 
 gate valve GV31 (or GV51).  It should then be possible to turn on the tank 
 ion gauge and HV, if time is of the essence and you want to resume use of 
 the separator. 
 Continue pumping on the cryo-pump until pressure drops to 100 mTorr 
 and when RV33 (or RV53) is closed the pressure rise is slow enough that 
 CP33 (or CP53) can be turned on before pressure goes above the 150 mTorr 
 interlock limit. 
 The temperatures of the CPs and the pressure at the associated CGs should 
 drop.  There should be a "rrr ... rrr ... rrr ..." sound from the 
 cold-heads at the ED tanks; if there isn't, try turning the power off and 
 on at the compressor (inside the fence, by the roughing and backing pumps) 
 and re-starting CP33, CP53.  If this doesn't work, consult an expert. 
 Go to the interlock Bypass page and turn Off the bypasses on RV33 
 and RV53 (yellow text should turn to black). 
 After pumping gunk from a cryopump, it often is seen that the 
 roughing pump RP21 can pump down only to approx. 150 mTorr instead of the 
 usual approx. 50 mTorr.  (This may be due to release of large amounts of 
 water from the cryos.)  Purge the roughing line and roughing pump: 
 close PV21 and vent the roughing line to nitrogen via VV21A (following the 
 instructions given above for venting a slit box volume); close VV21A and 
 pump out the roughing line through RV21.  If this improves the vacuum at 
 CG21B, a second purge may be useful.  If purging did not reduce the 
 pressure at CG21B, consult an expert. 
   Significant levels of radiation may be present when positron-emitting 
 radio-active beam is in use.  Approximately 60% of the beam stops at or 
 near the Charge slits following MD1 and 40% on the Mass slits.  Lead 
 shielding around the Charge and Mass slit boxes reduces the radiation to 
 acceptable levels at 1 m from the ion-optical axis of the separator. 
 A gated mesh fence prevents people from getting closer than 1 m to the 
 separator when radio-active beam is (or might be) delivered. 
 A number of gates allow access during maintenance periods or when stable 
 beam is being delivered from OLIS.  Each gate has a lock and key.  The key 
 can be removed only when the gate is locked; the key must be taken from 
 the gate and inserted in a "transfer panel" in order to release a key 
 which in turn is needed to get authorization for radio-active beam 
 delivery.  The transfer panel is located on a post to the north of the 
 DRAGON end detectors. 
The controller/power supplies for turbo pumps, ion gauges, motors, 
 etc. are mostly in 19-inch racks "near" the pump, gauge or motor being 
 controlled.  The important racks are: 
  
 Rack 7A - under the platform 
     
     MD1, MD2 NMR 
     HV4032A supply: XRAY1, XRAY2, BETA1, BETA2 PMT's 
     IG32 (south end) 
     TP32 (south end) 
     Rack 18A - East of ED1 tank
     
     ED1 HV interlock box 
     ED1 Anode HV supply 
     IG21, IG31 
     TP21, TP31 
     ED1 ion pump
     Rack 26A - Southwest of ED2 tank 
     
     ED2 HV interlock box 
     ED2 Anode HV supply 
     IG51, IG52 
     TP51, TP52 
     ED2 ion pump
     Rack 23A - on platform, North side
     
     EPICS VME crate 
     Stepping motor drivers
     Rack 23C - on platform, North side 
     
     Power supplies for SX1, SX2, SM1X, SM1Y, SM2X, SM2Y 
     MD1, MD2 NMR ref. voltage DACs
     Rack 23D - on platform, North side 
     
     Power supplies for Q1, Q2, Q3, Q4, Q5, SM0X, SM0Y
     Rack 24A - on platform, South side
     
     Power supplies for Q6, Q7, Q8, SM3X, SM3Y, SM4X, SM4Y
     Rack 24B - on platform, South side
     
     Power supplies for SX3, SX4, Q9, Q10 
      |