Minutes of the DRAGON meeting 13 November, 2007 Present: CR, GR, DH, CD, LB, PM, CV, UH, PB 0) Minutes of the previous meeting -- approved by default 1) Hardware: Ion pump committed suicide (PM) -- it burned through the HV feedthru and created a vacuum leak. Removed ion pump and blanked off; pumped down to low 10**-7. We need a database of loaned items (TUDA has 2, TITAN 1 of our pre-amps). 2) Solar Cells: Looking for grounding problems -- took apart and reassembled -- still problems but different (UH). Solar cells on common ground ``way back at the junction box'' (CR). CV fiddled with exterior cables and now things are different. Some channels have nothing; PAs saturated? We should go through every channel with scope (CR). Some questions about what source is being used. Taking a long time . . . . 3) Q1/Q2 adjustment . . . . so, Go ahead and move those quads (CR). Two adjustment points on top and one on the side (DH/PM). Extrapolate from student's (Michael) measurements to get how much we should move casing of the quad. Q1 relative to Q2 difference in height is certainly related to setting the y-slit C 3 mm high (DH). 3 to 1 magnification, so image is high ==> Q1 should be lowered (DH). Measure gap and check with TOTAL station. Height adjustment is most significant (DH). Make reversible changes (CR). Use alpha source to check change in image at charge slits as quad is moved (DH). Clearest signal is with strip detector DS of charge slits. Make sure that the source is in the right position and orientation (CR). JP had a technique with pumping tubes in place. DH will circulate a plan within the next 24 hours. CR: get started on this fairly quickly (will need DO's help). 4) A.O.B -Installation of super nano gun -- CR, CV, & DH [?] will help Kerthie get it installed. -All the EEC reports are in. 15 active proposals, 7 reports submitted, 2 new proposals.